Perusahaan ini terdaftar sebagai Free Member. Hindari melakukan pembayaran sebelum bertemu penjual atau melihat barang secara langsung. COD (Cash On Delivery) atau bertemu langsung dengan penjual merupakan metode transaksi aman yang kami sarankan.
I-V curve test for solar panel/module
Max. solar panel power (Pmax) search by auto-scan:
PROVA 200A: 60V, 6A (180W capability)
PROVA 210: 60V, 12A (500W capability)
PROVA 218: 85V, 8.5A (500W capability)
Best resolution for PROVA 200A: 1mV, 0.1mA
Best resolution for PROVA 210/218: 1mV, 1mA
Manual single point I-V test
Max. voltage (Vmaxp) at Pmax
Max. current (Imaxp) at Pmax
Voltage at open circuit (Vopen)
Current at short circuit (Ishort)
I-V curve with cursor to display each data point
Efficiency (%) calculation of solar panel
Solar panel area setting: 0.001 m2 ~ 9999 m2
Standard light source setting: 10 W/m2 ~ 1000 W/m2
Communicate with PC via USB cable
Min. power setting for alarm function
Built-in calendar clock
Built-in battery charging circuit
Rechargeable lithium battery and AC power adaptor
Memory size: 100 records
Sampling time of data logging for PROVA 210/218: 0 ~ 99 min.
Large LCD with backlight
Electrical Specifications: (PROVA 200A)
(23℃±5℃, four-wire measurement, the max. power limit is 180W)
DC Voltage Measurement:
Range
Resolution
Accuracy of Reading
0 ~ 6 V
0.001 V
±1% ± (1% of Vopen ± 9mV)
6 ~ 10 V
0.001 V
±1% ± (1% of Vopen ± 0.09V)
10 ~ 60 V
0.01 V
Vopen: open circuit voltage of solar cell or module.
DC Current Measurement:
Range
Resolution
Accuracy of Reading
0.01 ~ 0.6 A
0.1 mA
±1% ± (1% of Ishort ± 0.9mA)
0.6 ~ 1 A
0.1 mA
±1% ± (1% of Ishort ± 9mA)
1 ~ 6 A
1 mA
Ishort: short circuit current of solar cell or module.
Internal Resistance at Ishort: 0.05 Ohm.
DC Current Simulation*
Range
Resolution
Accuracy of Reading
0.01 ~ 1 A
0.1 mA
±1% ± 0.9mA
1 ~ 6 A
1 mA
±1% ± 9mA
* If current is greater than 6A, test (Auto-Scan, Scan, or Test) can not be performed.
* Max. duration of simulation is 9.999 seconds if power is less than 100W.
* Duration of simulation is 10m seconds if power is greater than 100W.
Electrical Specifications: (PROVA 210/218)
(23℃±5℃, 4-wire measurement, max. power limit 500W)
DC Voltage Measurement:
Range
Resolution
Accuracy
0 ~ 10 V
0.001 V
±1% ± (1% of Vopen ± 0.1V)
10 ~ 60 V (PROVA 210)
10 ~ 85 V (PROVA 218)
0.01 V
Vopen: open circuit voltage of solar cell or module.
DC Current Measurement:
Range
Resolution
Accuracy
0.01 ~ 10 A (PROVA 210)
0.01 ~ 8.5 A (PROVA 218)
1 mA
±1% ± (1% of Ishort ± 9mA)
10 ~ 12 A (PROVA 210)
10 mA
±1% ± (1% of Ishort ± 0.09A)
Ishort: short circuit current of solar cell or module.
DC Current Simulation
Range
Resolution
Accuracy
0.01 ~ 10 A (PROVA 210)
0.01 ~ 8.5A (PROVA 218)
1 mA
±1% ± 9mA
10 ~ 12 A (PROVA 210)
10 mA
±1% ± 0.09A
General Specifications:
AC Power Adaptor:
AC 100V~240V input
DC 15V / 1~3A output
Data Logging Memory Size:
100 records
Dimension:
257(L) x 155(W) x 57(H) mm
10.1″ (L) x 6.1″ (W) x 2.2″ (H)
Weight:
1160g / 40.9oz
(Batteries included)
Operation Environment:
5℃ ~ 50℃, 85% RH
Storage Environment:
-20℃ ~ 60℃, 75% RH
Temperature Coefficient:
0.1% of full scale / ℃
(<18℃ or >28℃)
Accessories:
User manual x 1;
Rechargeable lithium battery (PROVA 200A: 1600mAh; PROVA 210/218: 3400mAh) x 1;
AC power adaptor x 1;
Optical USB cable x 1;
Software CD x 1;
Software manual x 1;
Carrying bag x 1;
4-Wire to 2-Wire connector x 1 set;
Kelvin clips (2 clips) x 1 set (PROVA 200A: 6A max; PROVA 210/218: 12A max)
testing instrument